The ILX Lightwave LPA-9072B Laser Diode Pararneter Analyzer tests laser diodes/LEDs operating from 0 to 500. It can generate L/I, V/I, L/lm and dL/dl parametric data with 16-bit resolution in 10 seconds or less. Up to 10 sets of data can be compiled and presented in both graphical and tabular form. Data can be shared with other applications or easily output to a high-resolution printer. Specifications. Current Output; Maximum Current: 200 mA / 500 mA. Current Limit Range: 0 to 202 mA / 0 to 505 mA. Current Limit Resolution:1 mA on 200 mA range / 2 mA on 500 mA range. Set Point Resolution: 16 bit (4 µA / 10 µA). Set Point Accuracy (% of FS): ±0.05%. Compliance Voltage: 6.5V. Forward Voltage Measurement; Range: 0 to 7.0 V. Accuracy: ±2 mV. Photodetector Amplifier; PD Current Range: 10 nA to 50 mA. Accuracy (% of FS): ±5% (10 nA to 500 nA), ±1 % (500 nA to 50 mA). Short Term Repeatability @ 25 C: ±0.5% (10 nA to 500 nA), ±1 % (500 nA to 50 mA). Back Facet Monitor; Range: 1 µA to 5,000 µA. Accuracy (% of FS): ±0.05%. Reverse Bias: 0 to -5 V, adjustable. Stability (24 hr period): ±0.02%. Temperature Control; Range: -99.9°C to + 199.9°C. Resolution: 0.1°C. Thermistor Set Point Accuracy: ±0.2°C from 0°C to +50°C. TE Control: Bipolar, constant current source.
|